Test Bench for Heat Meter/ Water Meter
Size from DN15-DN25
Flow rate test/ Heat energy test/ Cooling energy test/ Temperature sensors test/
Water meter calculator test/ Heat meter calculator test
Test bench connects to meter via M-BUS, RS-485, Red Infrared
Design Fully Automatic or Semi-Automatic
PC data administration and download
Main Technical Specifications Nominal diameter(mm) 15-25 Main pipe(mm) 25 Checked meter caliber(mm) 15/20/25 Measuring method Static mass method Flow range (m³/h) 0.004~8 Pipeline through pipe Electronic weighing(KG) 30 Mettler-Toledo (optional) Electronic weighing(kg) 150 Mettler-Toledo (optional) Flow stabilizer (m³) 0.15 Water pump Wilo Operating pressure(MPa) 0.2-0.5 Device uncertainty(%) K=2 0.2 Working temperature(℃) 5-50℃ Working medium Clean water Device accuracy level 0.2 Number of meters DN15:12 PCS
DN20:10 PCS
DN25:8 PCS
Test Bench Components
Number | Name | Quantity | Remarks |
1 | Stainless steel flow detection table | 1PCS |
|
2 | Ultrasonic indicator flow meter | 5PCS | DN15*3 DN 20*1 DN 25*1 |
3 | Air operated valve | 10PCS | Water flow switch control valve: DN-15 * 3 sets; DN-20 * 1 set; DN-25 * 1 set |
4 | Manual ball valve | 9PCS | DN-15*4 DN-20 *1 DN-25*2 |
5 | Water pump | 1PCS | 2.2KW Horizontal pump WILMHI805 |
6 | Electronic scale | 2PCS | Toledo 150KG 1PCS number of verification scale intervals n=3000; |
7 | Weigh tank | 2PCS | The large barrel can weigh 140 kg |
8 | Surge pump | 1PCS | 0.15 cubic meters |
9 | Water tank | 1PCS | 1.2m*0.6m*0.55m 350L |
10 | Refrigerator | 1PCS | Air-cooled chiller AGR-05IA |
11 | IPC | 1PCS | (TARDETECH) TD-IPC-7101 |
12 | Touchscreen monitor | 1PCS | DELL |
13 | Distribution box | 1PCS | 380V,21KW |
14 | Control system | 1PCS |
|
15 | Air compressor | 1PCS | 220V ,0.1m3/min,0.6MPa |
16 | Heating pipe | 1PCS | 15KW |
17 | Thermostatic bath | 1PCS | 220V, 3KW, double ponds |
18 | Second-class standard platinum resistance | 2PCS |
|
19 | Normal thermometer | 1PCS | HY250A |